7-19 
Cause  FRU  Solution 
0x00000040 
Balance gas limb 
leakage 
EFCS assembly 
Restart the machine. If the problem 
persists, replace the EFCS assembly 
0x00000080 
error upon 
EFCS assembly 
Restart the machine. If the problem 
persists, replace the EFCS control board 
0x00000200 
Air three-way valve 
error 
EFCS assembly 
Restart the machine. If the problem 
persists, replace the EFCS assembly 
0x00000400 
N2O three-way 
valve error 
EFCS assembly 
Restart the machine. If the problem 
persists, replace the EFCS assembly 
0x00000800 
O2 two -way valve 
error 
EFCS assembly 
Restart the machine. If the problem 
persists, replace the EFCS assembly 
 
7.3.4.2 BFCS Error Codes 
Cause  FRU  Solution 
0x00000001 
NO two-way valve 
error 
EFCS assembly 
Restart the machine. If the problem 
persists, replace the EFCS assembly 
0x00000002 
Mechanical door 
status error 
BFCS door 
BFCS switch 
control board 
1. Open the BFCS door, check whether 
the magnet inside the BFCS door falls 
off. If yes, replace the BFCS door (refer 
to 
11.1.5). 
2. If the magnet inside the BFCS door 
does not fall off, replace the BFCS 
switch control board (refer to 11.1.5). 
0x00000004 
Electromagnet 1 
power-on failure 
Power board 
Electromagnet 
1 
EFCS control 
board 
1. If a power supply alarm exists, clear 
the alarm (refer to 7.3.3). 
2. Perform steps 1, 3, 6, 7, and 8 in 
section 4.6.5.2 to identify the faulty 
electromagnet, and then replace the 
faulty electromagnet (refer to 11.1.5). 
3. If the problem persists, replace the 
EFCS control board (refer to 11.3.1). 
0x00000008 
Electromagnet 1 
power-off failure 
Electromagnet 
EFCS control 
board 
1. Perform steps 1, 3, 6, 7, and 8 in 
section 4.6.5.2 to identify the faulty 
electromagnet, and then replace the 
faulty electromagnet (refer to 
11.1.5). 
2. If the problem persists, replace the 
EFCS control board (refer to 11.3.1). 
0x00000010 
Electromagnet 2 
power-on failure 
Power board 
Electromagnet 
2 
EFCS control 
board 
1. If a power supply alarm exists, clear 
the alarm (refer to 7.3.3). 
2. Perform steps 1, 3, 6, 7, and 8 in 
section 4.6.5.2 to identify the faulty 
electromagnet, and then replace the 
faulty electromagnet (refer to 11.1.5).