· Avoid installing the machine closer to the instrument which generates vibration.
· It is recommended that the installation place be on the first floor of the facility.
· Avoid installation on a facility close to the busy street with much traffic.
· Avoid installation on a facility where cranes or vibration hardness testing
machines are installed.
(The vibration may be absorbed by the hardness hardness testing machine, and it
can adversely affect the test result.)
Test force no more than 9.807×10
-3
N(1gf):
Vibration level: 0.2 Gal or less at 5Hz or lower
(The vibration higher than 5Hz is recommended to be damped using a vibration
isolating stand.)
Test force 9.807mN(1gf) to no more than 98.07mN(10gf):
Vibration level: 0.5 Gal or less at 5Hz or lower
(The vibration higher than 5Hz is recommended to be damped using a vibration
isolating stand.)
Test force 98.07mN(10gf) to no more than 1961mN(200gf):
Vibration level: 1 Gal or less at 5Hz or lower
Test force of 1961mN(200gf) or more
Vibration level: 5 Gal or less at 5Hz or lower
If exceeding the above condition, take the anti-vibration measure. For the anti-vibration
measure, please inquire separately.