No. 99MBG127A
5-4
3. Select the specimen and conversion scale, and then press [OK].
4. To exit conversion setting
Pressing restores the previous settings, and returns to the previous screen.
Pressing returns to the "MEASUREMENT" screen.
5.2.2 OK/NG judgment setting (JUDGEMENT)
When finalizing hardness measurement, the [OK] will be displayed if the test results between the
upper and lower limit. If the test result exceeds the hardness upper limit, the [+NG] will be
displayed, and if it falls below the hardness lower limit, the [-NG] will be displayed.
CAUTION
1. Judgment range of the hardness value: 0 to 9999
2. When setting a same value in upper limit and lower limit, the OK/NG judgment does
not function.
3. If trying to set the higher value than the upper limit to the lower limit, or vice versa, an
error message will be displayed. In this case, follow its instructions.
1. [Press the [JUDGEMENT] button. ⇒The "JUDGEMENT" screen is displayed.
Selection of
conversion scale
Selection of
specimen