5. Operation and various settings
No. 99MBG127A
5-17
5.4.1 Measuring points setting (NUMBER)
It sets the number of data to be statistically processed.
When statistical analysis and editing is not required, set it [OFF].
Important
1. The 256 points as fixed number will be measured when the statistical analysis is set
to [OFF]. When it is set to [ON], the testing numbers can be set between 2 to 256
points. When it is set to [OFF], "256" is displayed in "Measuring points".
2. When the testing reaches 257th points, the oldest data will be deleted.
No. 1 2 3 4 5 256
HV 700 720 710 705 720 695
No. 1 2 3 4 255 256
HV 720 710 705 720 695 715
3. When the testing reaches the set number while the setting is [ON], all the test data
will be deleted, and start storing data counting again from 1st point. The number of
points will be counted in ascending sequence from [1].
<Ex. When setting "256" points>
No. 1 2 3 4 5 256
HV 700 720 710 705 720 695
No. 1 2 3 4 5 256
HV 702
Deleted
Deleted
257th point test data