LIFEPAK 500 automated external defibrillator Replacement Procedures
11-44
Removing Main PCB—Biphasic AED Page 3 of 3
7. Disconnect the four spade-lugs P2, P4, P13, P15, and P19 from Main PCB
J2, J4, J13, J15, and J19, respectively. Do not pull on the wires.
8. Disconnect P11 of the data cable from J11 of the Main PCB.
9. Disconnect P16 from J16 on the Main PCB near Relay at J15. Disconnect
P19 from J19.
10. Carefully lift out the Main PCB, place it in a static-protective bag, and set it
aside.
11. After removing the Main PCB, jump to the appropriate replacement
procedure:
– Replacing Battery Plate Wire Harness
– Removing Biphasic PCB—Biphasic AED
– Replacing Inductive Resistor—Biphasic AED
– Installing Main PCB—Biphasic AED