EGM-5 Operation Manual V. 1.04  74  support@ppsystems.com 
 
Injection Baseline Phase Menu 
Pulses on and off to indicate that the system is powered on (power 
status). 
Percentage of battery life remaining. 
2
 concentration in tubing (ppm). 
2
 concentration inside syringe (ppm). 
2
 
Indicates that the process is in the Baseline phase for the first 10 
readings. 
Time (seconds). It is fixed at 180 seconds. 
CO
2
 concentration in tubing (ppm). 
2
 
maximum. This is set to be ¼ of the Max CO
2
for the Graph setting and can be modified under Main > Settings > 
Settings 2 > Graph. (For example, when Max CO2 is set to 1000 ppm, 
this graph shows 250 ppm.) 
Stops the Injection process and returns to Injection – Settings (Step 2). 
Process data is not saved to the USB flash drive. 
Stops the Injection process and returns to Injection - Settings (Step 2). 
Injection Process results are saved to the USB flash drive. 
 
If the baseline (B) does not drop to either 0 or a very small number (typically 1-3 ppm), then there is either 
a leak in the system, or the system has not adequately “zeroed”. Try aborting the process, wait 2-3 
minutes, and try again. If the problem persists, check the Injection setup for leaks and ensure that the 
absorber column is properly seated and contains fresh soda lime. 
Injection Step 6 – Injection Phase 
During the Injection Phase, the syringe should be slowly and steadily injected into the airstream through 
the septum. The injection rate should not exceed 3 ml/second to avoid over-pressurization of the system 
(in other words, it should take the user at least 3 seconds to inject a 10 ml syringe).