EGM-5 Operation Manual V. 1.04 117 support@ppsystems.com
Error Messages
Error and status messages are displayed in the Status Box during system operation. The following table
highlights the most common messages that you may see.
A/D values from the last Zero are too low (<2500).
2
2
setting (250 ppm by
default).
2
is greater than the high CO
2
setting (2000 ppm by
default).
Analyzer temperature < 50
o
C.
Analyzer temperature > 60
o
C.
Sample flow rate is less than 50 cc/min.
IRGA unable to perform a write operation. If error persists, call PP
Systems technical support.
IRGA read/write EEPROM CRC incorrect. If error persists, call
PP Systems technical support.
IRGA setting error. Value is out of range. If error persists, call PP
Systems technical support.
Reached Process CO
2
limit
Reached Process time limit
The SR measurement is Non Linear
The process has been terminated
The Absorber column material has been exhausted and needs
replacing
Zero will occur in 2 minutes
Zero will occur in 1 minutes
Could not successfully write the IRGA DigiPot
Host only. First character is not recognized as a command.
Host only. First character is a recognized command but the next
character is invalid. Example: S1,1<CR> (as opposed to
Host only. Command is correct but parameter is not valid.
Example: S,20,5. The value of 20 is an invalid parameter.
Host Only. Value entered is below the valid range of values for
that parameter.
Host Only. Value entered is above the valid range of values for
that parameter.
Flash drive error. Command failed – Filename or directory name
not found.
Flash drive error. Bad command – Command not recognized.
Flash drive full – No free space on disk.
Flash drive error. Invalid – Attempt to open a directory for reading
or writing. Attempt to change currently selected directory to a file.
Flash drive error. Read only – Attempt to open a read only file for
writing.
Flash drive error. File open – A file is currently open for writing
and must be closed before this command can be executed.