EGM-5 Operation Manual V. 1.04 117 support@ppsystems.com
Error Messages
Error and status messages are displayed in the Status Box during system operation. The following table
highlights the most common messages that you may see.
Error Messages
Error
Code
Status Box
Message
Description
10
Low Zero
A/D values from the last Zero are too low (<2500).
11
Low Alarm
Measured CO
2
is less than the low CO
2
setting (250 ppm by
default).
12
High Alarm
Measured CO
2
is greater than the high CO
2
setting (2000 ppm by
default).
13
Under Temp
Analyzer temperature < 50
o
C.
14
Over Temp
Analyzer temperature > 60
o
C.
15
Low Flow
Sample flow rate is less than 50 cc/min.
16
IRGA Write
IRGA unable to perform a write operation. If error persists, call PP
Systems technical support.
17
IRGA CRC
IRGA read/write EEPROM CRC incorrect. If error persists, call
PP Systems technical support.
18
IRGA Range
IRGA setting error. Value is out of range. If error persists, call PP
Systems technical support.
19
CO
2
limit
Reached Process CO
2
limit
20
Time limit
Reached Process time limit
21
Non Linear
The SR measurement is Non Linear
22
End Process
The process has been terminated
23
Absorber used
The Absorber column material has been exhausted and needs
replacing
24
Zero 2 min
Zero will occur in 2 minutes
25
Zero 1 min
Zero will occur in 1 minutes
26
IRGA DigiPot
Could not successfully write the IRGA DigiPot
30
Unknown CMD
Host only. First character is not recognized as a command.
31
Invalid Format
Host only. First character is a recognized command but the next
character is invalid. Example: S1,1<CR> (as opposed to
S,1,1<CR>.
32
Invalid Param
Host only. Command is correct but parameter is not valid.
Example: S,20,5. The value of 20 is an invalid parameter.
33
Low Limit
Host Only. Value entered is below the valid range of values for
that parameter.
34
High Limit
Host Only. Value entered is above the valid range of values for
that parameter.
50
No Memory
No flash drive detected.
51
Memory CF
Flash drive error. Command failed Filename or directory name
not found.
52
Bad Record
Flash drive error. Bad command Command not recognized.
53
Memory Full
Flash drive full No free space on disk.
54
Memory FI
Flash drive error. Invalid Attempt to open a directory for reading
or writing. Attempt to change currently selected directory to a file.
55
Memory RO
Flash drive error. Read only Attempt to open a read only file for
writing.
56
File Open
Flash drive error. File open A file is currently open for writing
and must be closed before this command can be executed.