EGM-5 Operation Manual V. 1.04 99 support@ppsystems.com
W
Auto
“W” String (Warm up)
The W string is transmitted every second during initial warmup, until the operating
temperature is reached. No “M” strings are sent during this time. The W string contains
the IRGA temperature, Battery Percent, and the Zero Column Percent.
Example:
W,55.0, 80, 59.47<CR>
In this example,
55.0 is the IRGA temperature (°C).
80 is the Battery % of Charge
59.47 is the Remaining Zero Column Capacity
Z
Auto
“Z” String (Zeroing)
The Z string indicates a zero in progress. During a zero operation, the EGM-5
transmits the Z string every 1 second, until the end of the zero operation. No “M”
strings are sent during this time. This command can be initiated by the user or by the
EGM-5 itself.
Example:
Z, 025<CR>
Z, 024<CR>
In this example, 025 and 024 represent the number of seconds remaining in the zero
operation.