EGM-5 Operation Manual V. 1.04 86 support@ppsystems.com
Hours Diagnostics
This screen displays the amount of time (in hours) that the controller and IRGA have operated in the
instrument’s lifetime.
Hours Diagnostics
Controller (hr)
Displays the total amount of time (in hours) that the controller has operated.
IRGA (hr)
Displays the total amount of time (in hours) that the IRGA has operated.
Back
Returns to the Diagnostics Menu.
Battery Diagnostics
This screen displays the percentage of battery remaining, the voltage and current of the battery, and the
time remaining before the battery is completely discharged.
Battery Diagnostics
Charge (%)
Displays the percentage of battery charge remaining.
Voltage (V)
Displays the battery voltage.
Current (A)
Displays the average battery current in Amps over the past 60 seconds. A
negative value indicates discharge current and a positive value indicates the
battery is being charged.
Time Left (M)
Displays the estimated remaining run time in minutes assuming the average
current does not change. A value of 65535 is displayed during battery charging.
Back
Returns to the Diagnostics Menu.