EGM-5 Operation Manual V. 1.04  81  support@ppsystems.com 
 
 
2 
2 
concentration (%) of the calibration reference gas. 
Entering a value will begin the Calibration Process 
The current O
2
 reading based on the Scale Factor 
The current Scale Factor being used 
Returns to the Calibration Menu. 
 
Connecting Calibration Gas to the EGM-5 
During calibration, the signal from the O
2
 sensor is referenced against a gas of known O
2
 concentration.  
For greater accuracy, reference gas mixtures should be accurate to at least +/- 1% and traceable to NIST 
standards.  
Follow these instructions to calibrate the EGM-5 O2 sensor: 
1.  To avoid excess pressure in the sample cell and damage to the analyzer, you must place a T-fitting 
between the reference gas source and the GAS IN port on the EGM-5. One end of the T-fitting should 
be left open in order to relieve excess pressure, as follows: 
 
 
O
2
 
GAS 
Cylinder
Regulator 
Valve 
To vent 
Gas 
~ 250ml/min 
Port