- 101 -
Ambient conditions recording interval [min]: parameter enabling you to specify record
frequency of sensor-registered data, and to determine how often the displayed
pictograms (providing information on ambient conditions state) are to be refreshed,
Internal sensor 1: enter this parameter to specify tolerance temperature values and
moisture analyzer temperature change rate.
Min temperature min moisture analyzer temperature, for
lower temperature values the thermometer pictogram is
red.
Max temperature max moisture analyzer temperature,
for higher temperature values the thermometer pictogram is
red.
Temperature delta t/hmaximum moisture analyzer
temperature change rate, for higher change rate the
moisture analyzer displays blinking red thermometer
pictogram.
In order to enable/disable parameter visibility go to service menu.
Ambient conditions module: parameters group which has been designed to enable you
to turn on readout of data recorded by a moisture analyzer-connected ambient
conditions module, and to specify tolerance values for this module.
Ambient conditions module - parameter specifying
module’s working mode. Available options: Nonereadout
of ambient condition module's indications turned off;
Recordreadout and record of indications into module's
database turned on; Record and alert readout and
record of indications into module's database turned on,
option of display of warnings informing on ambient
conditions change turned on (the warnings are displayed
on the basis of tolerance high and low thresholds).
Additional temperature sensorturning on readout of
indication of the additional temperature sensor connected
to the ambient conditions module.
Min temperature minimum temperature value, for lower
temperature values the thermometer pictogram is red.
Max temperature maximum temperature value, for
higher temperature values the thermometer pictogram is
red.
Temperature delta t/hmaximum temperature change
rate, for higher change rate the moisture analyzer displays
blinking red thermometer pictogram.
Min humidity minimum humidity value, for lower
humidity values the pictogram is red.
Max humidity maximum humidity value, for higher
humidity values the pictogram is red.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the RADWAG MA X2.IC.A Series and is the answer not in the manual?

RADWAG MA X2.IC.A Series Specifications

General IconGeneral
BrandRADWAG
ModelMA X2.IC.A Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals