Remote Control - Commands
R&S
®
NRP2
197User Manual 1173.9157.02 ─ 03
Command Preset and *RST value Remark
CALC<[1]..8>:LIM:FAIL? 0
Limit violation.
CALC<[1]..8>:LIM:FCO? 0
Number of limit violations is reset.
CALC<[1]..8>:LIM:UPP:DATA 0 W or 0 DB
Upper limit value for measure-
ments.
CALC<[1]..8>:LIM:UPP:STAT OFF
No monitoring function for upper
limit value.
CALC<[1]..8>:LIM:LOW:DATA 0 W or 0 DB
Lower limit value for measure-
ments.
CALC<[1]..8>:LIM:LOW:STAT OFF
No monitoring function for lower
limit value.
CALC<[1]..8>:MATH
For command description
see page 228(
The calculate block returns the
measured value of the sensor
assigned to the primary channel.
CALC<[1]..8>:REL 0 DBM or 0 DB
-
CALC<[1]..8>:REL:POW 0 DBM
-
CALC<[1]..8>:REL:RAT 0 DB
-
CALC<[1]..8>:REL:STAT OFF
Measurements are not converted
into a reference value.
DISP:ILL -
Display lighting remains
unchanged.
DISP:MESS OFF
Do not display messages.
DISP:MESS:TEXT -
No change.
DISP:MESS:TYPE MESS
Messages cannot be confirmed
manually.
DISP:SEL 1
First window selected.
DISP:SIZE NORM
Automatic window sizes.
DISP<[1]..4> ON | OFF
A window is opened automatically
for each connected sensor.
DISP<[1]..4>:AVAL NONE
No additional information on mea-
sured values displayed.
DISP<[1]..4>:FORM DIG
Digital result display.
DISP<[1]..4>:MET:LOW -90 DBM
-120 DB
Lower limit value for analog result
display.
DISP<[1]..4>:MET:UPP 70 DBM
60 DB
Upper limit value for analog result
display.
DISP<[1]..4>:NAME ""
Text to be displayed in result win-
dow.
DISP<[1]..4>:RES 0.01
Measurement accuracy and dis-
play accuracy.
DISP:SEL 1
Selection of first result window.
Common Commands to IEEE 488.2