Remote Control - Commands
R&S
®
NRP2
225User Manual 1173.9157.02 ─ 03
CALCulate<[1]..8>:LIMit:BEEP <beep>
Switches the acoustic warning for limit violations on or off.
Parameters:
<beep> ON | OFF
*RST: OFF
Usage: Setting only
CALCulate<[1]..8>:LIMit:CLEar[:IMMediate]
Resets the limit monitoring state and the internal counter for limit violations.
Usage:
Event
CALCulate<[1]..8>:LIMit:CLEar:AUTO <auto>
Automatically resets the limit monitoring state and the internal counter for limit violations
if any of the following events occurs:
INITiate[:IMMediate] is executed
INITiate:CONTinuous ON is executed
MEASure? or MEASure:...? query is executed.
READ? or READ:...? query is executed.
Parameters:
<auto> ON | OFF | ONCE
ON
automatic reset enabled
OFF
automatic reset disabled
ONCE
automatic reset is enabled only once for the next occurrence of
the above events
*RST: OFF
Usage: Event
CALCulate<[1]..8>:LIMit:FAIL?
Queries whether upper or lower limits have been exceeded. The status is reset (0 = no
limit violation) if one of the following events occurs:
The device is switched on
A reset is performed (*RST)
The CALCulate<[1]..8>:LIMit:CLEar command is sent.
Usage:
Query only
CALCulate (Configuration of Calculate Blocks)