Remote Control - Commands
R&S
®
NRP2
226User Manual 1173.9157.02 ─ 03
CALCulate<[1]..8>:LIMit:FCOunt?
Queries the number of limit violations that occurred. The counter is zeroed if one of the
following events occurs:
The device is switched on
A reset is performed (*RST)
The CALCulat<[1]..8>:LIMit:CLEar command is sent.
Usage: Query only
CALCulate<[1]..8>:LIMit:LOWer[:DATA] <lower_limit>
Sets a lower limit for the measured values. The R&S NRP2 can respond when the values
drop below this limit (see other CALCulate<[1]..8>:LIMit commands).
Parameters:
<lower_limit> float_value
The value range depends on the current output unit of the calcu-
late block's measured value, see table below.
The parameter unit is determined according to table 6-11 Output
unit of measured value of calculate block.
*RST: 0 DBM or 0 DB
Default unit: Default unit: see table referred to above.
Table 6-10: Value ranges, depending on the unit
Unit of mea-
sured value
W DBM DBUV DPCT DB O
Value range -9.99999E5
..
9.99999E5
-120
..
90
-13.0103
..
196.897
-100
..
999.99
-200
..
200
-1E20
..
1E20
CALCulate<[1]..8>:LIMit:LOWer[:DATA]:POWer <lower_limit>
Sets a lower limit for the measured power. The R&S NRP2 can respond when the values
drop below this limit (see other CALCulate<[1]..8>:LIMit commands).
Parameters:
<lower_limit> float_value; units: W | DBM | DBUV
Range: -120 DBM to 90 DBM
*RST: 0 DBM
Default unit: DBM
CALCulate<[1]..8>:LIMit:LOWer[:DATA]:RATio <ratio>
Sets a lower limit for the measured power ratios. The R&S NRP2 can respond when the
ratios drop below this limit (see other CALCulate<[1]..8>:LIMit commands).
CALCulate (Configuration of Calculate Blocks)