Remote Control - Commands
R&S
®
NRP2
227User Manual 1173.9157.02 ─ 03
Parameters:
<ratio> float_value; units: DB | DBCT | O
Range: -200 DB to 200 DB
*RST: 0 DB
Default unit: DB
CALCulate<[1]..8>:LIMit:LOWer:STATe <state>
Switches the monitoring function for theupper and lower limit on or off.
Parameters:
<state> ON | OFF
*RST: OFF
Usage: Setting only
CALCulate<[1]..8>:LIMit:STATe <state>
Switches the monitoring function for the upper limit on or off.
Parameters:
<state> ON | OFF
*RST: OFF
Usage: Setting only
CALCulate<[1]..8>:LIMit:UPPer[:DATA] <upper_limit>
Sets an upper limit for the measured values. The R&S NRP2 can respond when this limit
is exceeded (see other CALCulate<[1]..8>:LIMit commands).
Parameters:
<upper_limit> float_value
The value range depends on the current output unit of the calcu-
late block's measured value, as shown in table 6-10 Value ranges,
depending on the unit.
The parameter unit is determined according to table 6-11 Output
unit of measured value of calculate block.
*RST: 0 DBM or 0 DB
Default unit: see table referred to above.
CALCulate<[1]..8>:LIMit:UPPer[:DATA]:POWer <power>
Sets a upper limit for the measured power. The R&S NRP2 can respond when this limit
is exceeded (see other CALCulate[1]..8>:LIMit commands).
CALCulate (Configuration of Calculate Blocks)