Remote Control - Commands
R&S
®
NRP2
281User Manual 1173.9157.02 ─ 03
Parameters:
<rlevel> float_value
Range: depending on sensor
*RST: depending on sensor
Default unit: DBM
[SENSe<[1]..4>]:STATistics:TIME < time_interval>
Determines, together with the SENS:STAT:OFFS command, the time interval in which
the power for the statistical evaluation is measured. SENS:STAT:TIME determines the
length of the interval (see also [SENSe<[1]..4>][:POWer]:TGATe<[1]..4>:
TIME on page 274).
Parameters:
< time_interval> float_value
Range: depending on sensor
*RST: depending on sensor
Default unit: S
[SENSe<[1]..4>]:TIMing:EXCLude:STARt <exclude_start>
Sets a time that is to be excluded at the beginning of the integration period.
Integration period
EXCLude:STARt EXCLude:STOP
Power
Time
Fig. 6-15: Effect of commands SENS<[1]..4>:TIM:EXCL:STAR and :STOP
Parameters:
<exclude_start> float_value
Range: depending on sensor
*RST: depending on sensor
Default unit: S
[SENSe<[1]..4>]:TIMing:EXCLude:STOP <exclude_stop>
Sets a time that is to be excluded at the end of the integration period. See figure 6-15
Effect of commands SENS<[1]..4>:TIM:EXCL:STAR and :STOP.