URISYS 2400 2 Hardware
Installation
Additional items
Items that are necessary or optional for the installation and operation of the
URISYS 2400 analyzer, but are not supplied with the analyzer, are:
o Liquid waste tank (with or without an optional overflow sensor)
o Racks
o Sample tubes
o Washing solution
o URISYS 2400 cassette
o URISYS 2400 calibration strip
o External printer and standard printer cable (parallel/Centronics interface)
o Waste container cartons
o Serial interface (host) cable
o 1.44 MB diskette for storage of data
Installation
Read this manual carefully before installation, to ensure correct operation of
the analyzer.
If you have any questions after initial installation, contact Technical Support.
Basic requirements
Surface and space
measurements
The following surface and space requirements must be met for the use of the
URISYS 2400 analyzer, and the proper ventilation of the analyzer when it is in
use:
o 5 cm (2.0 inches) or more (rear)
o 20 cm (7.9 inches) or more (left side)
o 30 cm (11.8 inches) or more (right side)
Use a table that is able to support the mass of the analyzer, which is
approximately 85 kg (187 lbs). The table must be level. Any inclination must
be less than 1.5 degrees in every direction.
Only trained Roche Technical Support personnel, or similarly qualified personnel
supervised by authorized service agents of Roche Diagnostics, are qualified to
install the URISYS 2400 analyzer.
At least two persons must carry the analyzer, by holding the baseplates on the left
side and right side of the analyzer. Be careful not to hurt your hands or fingers when
putting the analyzer in place.
Follow the specified installation conditions carefully. Otherwise, inaccurate results,
or damage to the analyzer may occur.