Roche Diagnostics
Operator’s Manual · Version 1.0 E-9
URISYS 2400 8 Data alarms (flags)
Reset the analyzer
Reset the analyzer
The reset procedure is used as a troubleshooting tool.
a To reset the analyzer
1 Press <Menu> on the [Routine Monitor] screen. The [Menu] screen is
displayed.
2 Press <Maintenance>. The [Maintenance] screen is displayed.
3 Press <Reset>. The [Reset] window appears.
4 Press <Execute> to start the resetting.
During Reset, the analyzer clears unprocessed racks from the cross-feed line,
discards test strips still present on the TS transfer base, performs a rinsing
cycle of the liquid handling system and resets these modules to their home
position.
Start
If the analyzer does not switch on at all, check that:
o The main power is on.
o The power cable is connected properly.
Contact Technical Support if the analyzer still does not respond.