Roche Diagnostics
C-24 Operator’s Manual · Version 1.0
5 Daily operation URISYS 2400
Analysis and documentation of results
Edit result data
a To edit routine and STAT sample data
1 Search for the requested sample by using the sequence or ID number, or
the <Prev>/<Next> buttons.
2 Press <Edit Data>. The [Edit Data] screen is displayed.
3 Press “<” or “>” to select the test parameter to be edited.
4 Press “+” or “-” to increase or decrease the result.
5 Press <Save> when editing is completed. The [Confirmation] window
appears.
6 Press <Yes> to return to the [Edit Data] screen. Check if the changes are
correct.
7 Press <PgUp> to return to the [Routine Data Monitor] screen.
When result data are edited, they are printed with an “!” edit mark.
Print routine and STAT
sample data
This print option is available for batch printing. For more information on
printing in real time, see:
a To print routine and STAT sample data
1 Press <Type> on the [Routine Data Monitor] screen to select the type of
data to be displayed (routine or STAT).
2 Press <Media>. The [Routine Data Monitor Media] screen is displayed.
3 Enter the range of sequence numbers to be printed, using numbers
between 1 and 1000 for routine sample data, or between 1 and 200 for
STAT sample data. Enter 0 to print all sample data.
4 Press <Print>. The [Confirmation] window appears.
5 Press <Data Select> to select the category of results to be printed (All,
Abnormal, Sieve, Normal).
6 Press <OK> to start printing.
o
To edit sample data, the analyzer must be in Stand-By status
o
Data cannot be edited if a sample is short or empty, or if a test strip error
occurred, (see Table E-1 on page E-2 and “Instrument alarms
(messages)” on page E-12).
“To select real time printouts and specify the result data to be printed” on page C-7