Roche Diagnostics
C-26 Operators Manual · Version 1.0
5 Daily operation URISYS 2400
Analysis and documentation of results
Send result data to the
host computer
This option is available for batch sending.
To send data, the host connection must be activated.
a To send routine or STAT sample data to the host computer
1 Press <Type> on the [Routine Data Monitor] screen to select the type of
sample data to be sent (routine or STAT).
2 Press <Media>. The [Routine Data Monitor Media] screen is displayed.
3 Enter the range of sequence numbers to be sent, using numbers between 1
and 1000 for routine sample data, or between 1 and 200 for STAT sample
data. Enter 0 to send all sample data.
4 Press <Send>. The [Confirmation] window appears.
5 Press <OK>.
6 Press <Quit> to terminate the transfer of data at any time.
Write/read routine or
STAT sample data to/
from a diskette
This option allows for data to be saved on a diskette or for data to be read
from a diskette.
a To write/read routine or STAT data to/from a diskette
1 Press <Type> on the [Routine Data Monitor] screen to select the type of
sample data to be written/read (routine or STAT).
2 Press <Media>. The [Routine Data Monitor Media] screen is displayed.
3 Enter the range of sequence numbers to be written/read to/from a diskette,
using numbers between 1 and 1000 for routine sample data, or between 1
and 200 for STAT sample data. Enter 0 to write/read all sample data.
4 Press <FD R/W>. The [Confirmation] window appears.
5 Press <File Type> to select the format of the files to be written/read (ASCII
or binary).
6 Insert appropriate diskette.
To define the communication parameters on page B-26
To activate/deactivate the connection to the host computer on page C-7
To send sample data, the analyzer must be in Stand-By status.