Roche Diagnostics
B-12 Operator’s Manual · Version 1.0
3 Software URISYS 2400
Installation
Installation
The software supplied with the analyzer supports all functions of the
URISYS 2400 analyzer. Specific functions are password-protected to avoid
accidental modifications.
The English software is pre-installed and ready for use on delivery of the
analyzer.
In case re-installation of the software is necessary, proceed as follows.
a To install the software
1 Save the system parameters, test parameters, routine sample results, STAT
sample results, control results and the alarm trace onto a diskette as safety
backup.
2 Switch off the power.
3 Insert System Disk 1 into the diskette drive.
4 Switch on the power, located on the right side of the analyzer. The [Wait]
screen is displayed.
5 While the [Wait] screen is displayed, press both the lower left and lower
right corners of the LCD panel simultaneously. The [System Menu] screen
is displayed.
6 Press <Transmit Mode>.
7 Press <Tool Transmit>.
8 Follow the instructions on the screen to complete the installation.
9 If necessary, restore the system parameters, test parameters, routine sample
results, STAT samples results, and control results from the backup diskette.
“Setting the parameters” on page B-16
“Data of routine and STAT samples” on page C-21
“Control sample data” on page C-28
“Handle alarms” on page E-3