Operator’s Manual · Version 1.0 C-21
URISYS 2400 5 Daily operation
Analysis and documentation of results
Analysis and documentation of results
Using the <Data Monitor> menu, the operator can perform the following
tasks for routine, STAT, and control samples:
o Display the result data
o Search result data
o Enter sample ID number
o Edit the result data
o Print the result data
o Send the result data to the host computer
o Write/read the result data to/from a diskette
o Delete the result data
The result data can be displayed at any analyzer status – Operation, Sampling
Stop, or Stand-By. However, for documenting result data, the analyzer must
be in Stand-By status.
The following sections have been divided into two parts:
o Data of routine and STAT samples
o Data of control samples
Data of routine and STAT samples
The analyzer has a memory capacity of 1000 routine sample results and a
separate memory for 200 STAT sample results. The results of all test
parameters for a selected sequence number or sample ID number are
displayed on the screen.
If no results are available for a processed sequence number, a message, such as
“Sample empty”, is displayed on the screen and printed on the printout.