URISYS 2400 5 Daily operation
Analysis and documentation of results
6 Press <Concentr.> to change to the display of the results as reflectance
values of the same control sample, or <Reflectance> to return to the
concentration values.
Search for control
sample data
Use the sequence number and the level category of the controls to search for
data of control samples.
Results cannot be retrieved using the control name, lot number, date and
time, or operator ID.
a To search for control sample data using the sequence number
1 Press <Level> on the [Control Data Monitor] screen to select the level of
controls to be displayed.
2 Press <Seq. No.> to enter the requested sequence number.
3 The [Control Seq. No.] window appears.
4 Enter the sequence number, using numbers between 1 and 100.
5 Press <Enter>. The result data of the entered sequence number are
displayed.
6 Press <Prev> or <Next> to display previous or subsequent control samples
data.
7 Press <Concentr.> or <Reflectance> to select concentration data or
reflectance data.
Figure C-9 Control Data Monitor screen (reflectance values)