Roche Diagnostics
C-12 Operator’s Manual · Version 1.0
5 Daily operation URISYS 2400
Remove processed samples
a To continue analysis of samples
1 Place the next racks directly into the rack input line.
2 Press <Start> on the [Routine Monitor] screen. The analyzer continues
sampling and measurement without an intermediate reset.
a To interrupt and restart analysis
1 Press <Stop> on the [Routine Monitor] screen. The confirmation window
appears.
2 Press <OK>. The status of the analyzer changes to Stand-By via Stop
status.
Automatic termination
If there are no more racks on the rack input line, sampling is automatically
terminated. The analyzer moves to Sampling Stop status. The processing of
test strips continues until all samples that have been pipetted onto test strips
have been measured.
Termination due to an alarm
Various alarms can terminate the processing of samples. For more
information on alarms, see:
Remove processed samples
Processed sample racks are automatically transferred to the tray on the rack
output line. If the rack tray is full, an alarm occurs (Rack Full). The analyzer
moves to Sampling Stop status.
a To remove processed samples
1 Remove the tray from the rack output line.
2 Remove the racks from the tray.
3 Place the empty tray on the rack output line.
When <Start> is pressed again, the subsequent reset clears the racks still present in
the cross-feed line without sampling them, and removes test strips still present on
the TS transfer base without measuring them.
“Instrument alarms (messages)” on page E-12