B-18 Operators Manual · Version 1.0
3 Software URISYS 2400
Installation
a To set the criteria for defining sieve and abnormal values
1 Press <Sieve & Abnormal> on the [Test Parameters] screen. The [Sieve &
Abnormal] screen is displayed listing the sieve criteria for each test.
2 Press <Select> to change the display of the items from sieve to abnormal
criteria and back.
3 Press <Set> to change the criteria. The [Password] window appears.
4 Enter the 4-digit password using the keyboard on the screen.
5 Press <Enter>. The [Sieve & Abnormal Setting] screen is displayed with
both the sieve and abnormal values listed for each test parameter,
beginning with specific gravity (SG).
6 Press <Test> to select another test parameter.
7 Press < or > to select sieve or abnormal settings.
8 Press <+> or <-> to select the desired concentration range.
9 Press <Enter> after each selection to accept the chosen values.
10 Press <PgUp> to return to the [Sieve & Abnormal] screen and check if the
settings correspond to your requirements.
11 Press <PgUp> to return to the [Test Parameters] screen.
Print order The operator can define the order of the test parameters on the screen and on
the result printout, in order from 1 to 12. If an item is not to be printed, enter
zero.
Figure B-10 Screen showing Print Order settings and options