URISYS 2400 2 Hardware
Installation
Assemble the parts
Set up the water supply
tank
The external water supply tank provides water to the analyzer for washing.
The tank has a volume of approximately 5 liters (5.3 quarts), which allows for
the measurement of about 1000 samples.
a To set up the water supply tank
1 Fill the tank with deionized water.
2 Connect the cable of the liquid level detector to the water level connection
on the analyzer.
3 Connect the water supply tube with filter to the water inlet connection on
the analyzer.
4 Press the water tube into the slit of the rubber stopper that is on top of the
liquid level detector, so the filter, at the end of the water supply tube will be
long enough to reach the bottom of the tank.
5 Put both the liquid level detector and the water supply tube with filter
through the opening of the tank, so the filter touches the bottom of the
tank, while the level detector with rubber stopper align vertically into the
tank.
Figure B-3 Connecting the water supply tank and liquid waste tank to the analyzer
A Water inlet connection
B Liquid waste tube connection
C Connector for water level detector
D Connector for liquid waste level detector
A
C
B
D
If the water is contaminated by the growth of bacteria or algae, it may result in faulty
liquid detection and rinsing in the pipetting mechanism and rinse circuits of the
analyzer.