5 Daily operation URISYS 2400
Switch off the analyzer
Switch off the analyzer
The URISYS 2400 analyzer must be in Stand-By status when it is switched off.
Never switch off the analyzer when it is in any other status. Otherwise, the
final rinsing procedure may be interrupted, or a sample may remain inside the
pipetting system overnight, which can cause a malfunction when the analyzer
is started again.
In addition, perform the required daily maintenance procedures.
1 If the analyzer is in S.Stop status as the result of an alarm, resolve the
alarm, and perform a reset procedure before switching off the power.
2 Switch off the power to the analyzer.