B-10 Operators Manual · Version 1.0
2 Hardware URISYS 2400
Installation
Environmental conditions
The URISYS 2400 analyzer must be used in an environment that meets the
following conditions:
o Free from excessive dust
o Well-ventilated area
o Not exposed to direct sunlight
o Floor that is level (with an inclination of less than 1.5 degrees in every
direction) and firm
o Table that can support the mass of the instrument, which is approximately
85 kg (187 lbs)
o Ambient temperature ranging from 15-30°C (59-86°F), with an acceptable
variation of ±2°C (±3°F) during measurement
o Relative humidity ranging from 20% to 80%, without moisture
condensation
o Free from vibrations
o Availability of a 3-pin power supply socket within 3 m (9.8 ft) of the
analyzer
o Free from abrupt mains voltage fluctuation (±10 V)
o Well-distanced from a machine generating a high frequency voltage (for
example, a centrifuge)
o Free from electromagnetic wave interference