EasyManua.ls Logo

Roche URISYS 2400 - Print Control Sample Data; Presentation of Result Data on the Printout

Roche URISYS 2400
166 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Roche Diagnostics
C-30 Operators Manual · Version 1.0
5 Daily operation URISYS 2400
Analysis and documentation of results
Print control sample data This option allows for batch printing of control sample data of the selected
level.
a To print control sample data
1 Press <Level> on the [Control Data Monitor] screen to select the level of
controls.
2 Select <Concentration> or <Reflectance>.
3 Press <Media>. The [Control Data Monitor Media] screen is displayed.
4 Enter the range of sequence numbers to be printed between 1 and 100.
Enter 0 to print all data.
5 Press <Print>. The [Confirmation] window appears.
6 Press <OK>.
7 Press <Stop> to terminate printing at any time.
Presentation of result
data on the printout
The print order of the test parameters results in concentration values is
identical to the order selected for routine samples.
The printout includes the following:
o Headline
o Date and time of the printout
o The sequence number of the sample (with a leading C1 for control samples
in position 1 of the rack, C2 for position 2 and C3 for position 3)
o Rack number and the position of the sample in the rack
o Operator ID
o Date and time of measurement
o Control name
o Control lot number
o Lot number and loading date and time for the URISYS 2400 cassette
o Test parameter results
o Flags
To print control sample data, the analyzer must be in Stand-By status.
Print order on page B-18
Table E-1 on page E-2

Table of Contents

Related product manuals