Roche Diagnostics
E-2 Operator’s Manual · Version 1.0
8 Data alarms (flags) URISYS 2400
Flags on result data printout
Flags on result data printout
The result data that are printed must be checked for flags. There can be
different flags next to the data, in order to bring some of the values to the
attention of the operator. Flags on result data printout are outlined below.
Flag Data alarm Explanation
Appearance
on printout
S Sieve Test parameter result corresponds
to sieve criteria settings
After the value
* Abnormal Test parameter result corresponds
to abnormal criteria settings
Before the
value
O Specific gravity
(SG) out of range
Result is greater than 1.050 or less
than 1.000
Instead of SG
value
I SG value invalid Refraction position is not definite Instead of SG
value
T SG error Hardware error Instead of SG
value
# Changed range
value
Initial range table is changed After the value
! Edit Data have been edited After the value
Table E-1 Flags on result data printout
Flag Sample alarm Explanation
Appearance
on printout
E Sample empty No sample detected Instead of
value
N Sample short Sample volume below limit Instead of
value
R Test strip Test strip error Instead of
value
Table E-2 Sample alarm flags