Roche Diagnostics
F-2 Operator’s Manual · Version 1.0
10 Technical specifications URISYS 2400
Technical specifications
Technical specifications
Measurement methods
System
description
Interface: 1 serial RS 232 C interface
1 parallel printer port
Display: Color LCD touch screen
Printer: External printer
(for example, Fujitsu-DL 3700 Pro)
Dimensions Height:
Width:
Depth:
Weight:
67 cm (26.4 in)
53 cm (21 in)
65 cm (26 in)
85 kg (187 lbs)
Power
supply
Supply voltage:
Power consumption:
Over-voltage category:
Protection class:
Pollution degree:
Fuses:
100-240 V AC (50-60 Hz)
0.2 kVA
II
1
2
Littelfuse Inc. Type 218315, T 3.15A, 250V
(2 pcs.)
Heat
dissipation
700 Kj (663 BTU)
Operating
conditions
Temperature: Operating: 15 - 30°C (59 - 86°F)
Storage and
transport:
-20 - 60°C (-4 - 140°F)
Stand by: 5-40°C (41 - 104°F)
Test strip parameters Type:
Light source:
Wavelengths:
Measuring head:
Reflectance photometry
LEDs (light-emitting diodes)
470 nm, 555 nm, 620 nm
1 measuring head with 3x3 LEDs
Clarity (CLA) Type:
Light source:
Wavelength:
Detector:
Turbidimetry
LED
660 nm
Photodiode
Specific gravity (SG) Type:
Light source:
Wavelength:
Detector:
Refractometry
LED
650 nm
CCD element
Processing Incubation time:
Cycle time:
60 sec
15 sec