Roche Diagnostics
Operator’s Manual · Version 1.0 F-3
URISYS 2400 10 Technical specifications
Barcode specifications
Barcode specifications
Item Specification
Reading method Scanning via LED and CCD sensor
Barcode symbol used NW7 (Codabar), Code 39, ITF
(Interleaved 2 of 5), Code 128
Check digit Usable
NW7 Modulus 16
Code 39 Modulus 43
ITF Modulus 10/weight 3
Code 128 Modulus 10/weight 3
Number of ID digits
NW7 3 to 13 digits (without check digit);
3 to 13 digits + 1 digit (check digit)
Code 39 3 to 13 digits (without check digit);
3 to 13 digits + 1 digit (check digit)
ITF 4 to 12 digits (without check digit);
3 to 13 digits + 1 digit (check digit)
Code 128 4 to 12 digits + 2 digits (check digit)
Characters usable
NW7 0 to 9, -, /, $, +
Code 39 0 to 9, space, A to Z, -, ., /, $, %, +
ITF 0 to 9
Code 128 Alphanumerical function characters
and communication characters are
excluded.
If using an ITF barcode system, Roche/Hitachi recommends its use only with a
check digit function.