Roche Diagnostics
Operator’s Manual · Version 1.0 B-25
URISYS 2400 3 Software
Installation
Sample barcode
a To select the type of barcode
1 Press <Sample Barcode> on the [System Parameters] screen. The [Sample
Barcode] screen is displayed.
2 Press <Sample Barcode> to enable or disable the use of a barcode.
3 Press <Code39 C/D>, <NW7 C/D>, and <ITF C/D> to enable or disable
the use of a check digit function for each type of barcode. A [Warning]
window will alert the operator of the increased probability for misreads
when check digits are disabled.
4 Press <PgUp> to return to the [System Parameters] screen.
Printer
a To define printer settings
1 Press <Printer> on the [System Parameters] screen. The [Printer] screen is
displayed.
2 Press <Paper Length> to define the paper size.
3 Press <Samples/Page> to define the number of samples to be printed per
page.
4 Press <Headline edit>. The [Headline edit] screen is displayed.
5 Enter the new headline (up to 26 characters per line).
6 Press <Enter> at the end of each line.
7 Press <PgUp> twice to return to the [System Parameters] screen.
If sample barcode is enabled, any sample with no barcode label or an unreadable
barcode, results in an alarm message. In this case, samples are still tested. If sample
barcode is disabled, no sample barcodes are read.
If using an ITF barcode type, Roche/Hitachi recommends its use only with a check
digit function.