EasyManua.ls Logo

Roche URISYS 2400 - Units; System Parameters

Roche URISYS 2400
166 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Roche Diagnostics
B-24 Operators Manual · Version 1.0
3 Software URISYS 2400
Installation
Units
a To select the concentration units
The operator can select the units for reporting results (conventional or SI).
1 Press <Units> on the [Test Parameters] screen. The [Units] screen is
displayed. The units currently in use are displayed.
2 Press <Units> to change the type of units to be displayed on the screen and
on the printouts, and to be sent to the host computer. The [Confirmation]
window appears.
3 Press <OK> to change units.
System parameters
The operator can select six system parameter settings. They are:
o Sample barcode
o Printer (includes the paper length, the number of samples printed per
page, and the page heading)
o Communication
o Date and time
o Password
o Registration of the wash rack
a To set system parameters
1 Press <System Parameters> on the [Parameters] screen. The [System
Parameters] screen is displayed.
When the units are changed, results are reported in the newly chosen units.
Therefore, it is important to set units before starting measurements. The change of
units also applies when re-printing previously measured results.

Table of Contents

Related product manuals