SME Analog Modulations
1038.6002.02 2.61 E-13
SYNC signal
VIDEO signal
RF signal
WIDTH
TRIGGER DELAY
WIDTH
Pulse input
Double Pulse Delay
Fig. 2-30 Signal example 2: double pulse, TRIGGER MODE = EXT, SLOPE = POS
Note: The minimum period length depends on parameters WIDTH and PULSE DELAY. To avoid a
settings conflict the following has to be true:
PERIOD
≥
1,1 × (WIDTH + PULSE DELAY) + 30 ns
For double pulse the following has to be true:
DOUBLE PULSE DELAY
≥
WIDTH + 40 ns
Menu MODULATION-PULSE offers access to settings for pulse modulation and to the pulse generator.
If only option SM-B3, pulse modulator, is fitted, only the first 3 lines are displayed in the setting menu.
Menu selection: MODULATION - PULSE
FREQUENCY
LEVEL
MODULATION
DIGITAL MOD
LF OUTPUT
SWEEP
LIST
MEM SEQ
UTILITIES
HELP
AM
FM
PM
PULSE
FREQ
100. 000 000 0
MHz LEVEL
- 30.0
dBm
SOURCE
OFF EXT PULSE-GEN
POLARITY NORM INV
EXT IMPEDANCE 50 10 k
---------------PULSE GENERATOR----------------
PERIODE 10.00 µs
WIDTH 1.00 µs
PULSE DELAY 1.00 µs
DOUBLE PULSE STATE OFF ON
TRIGGER MODE AUTO EXT
EXT TRIG SLOPE POS NEG
Fig. 2-31 Menu MODULATION-PULSE (preset setting), fitted with option SM-B3, pulse modulator,
and option SM-B4, pulse generator.