EasyManua.ls Logo

Rohde & Schwarz SME02 - AM Distortion Factor; Residual Phim with am

Rohde & Schwarz SME02
462 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
SME Test Procedure
1038.6002.02 5.21 E-13
5.2.18 AM Distortion Factor
Test equipment
Test system 5.1.1.1.
Test system 5.1.1.6 (SME06)
Measurement
À Settings at the SME
- Level 0 dBm
- Select LFGEN1 in the MODULATION/AM/AM SOURCE INT menu,
- Modulation depth 30%
- Modulation frequency 1 kHz.
À Vary the carrier frequency from 1 MHz to 1500 MHz (2200/
3000/6000 MHz with SME03E/03/06). For recommended setting
values, see table 5-2.
À Read off the distortion factor at the modulation analyzer.
Distortion factor ..................................................................max. 1%
À Repeat measurement with an AM of 80%.
Distortion factor ..................................................................max. 2%
5.2.19 Residual PhiM with AM
Test equipment Test system 5.1.1.1.
Test system 5.1.1.6 (SME06)
Measurement
À Settings at the SME
- Various test frequencies (for recommended setting values, see
table 5-2)
- Level 0 dBm
- Select LFGEN1 in the MODULATION/AM/AM SOURCE INT
menu,
- Modulation depth 30%
- Modulation frequency 1 kHz.
À Measure the phase modulation resulting by means of a 23-kHz
lowpass filter and peak weighting at the modulation analyzer.
Unwanted modulation with frequencies
3 GHz...........max. 0.1 rad
Unwanted modulation with frequencies > 3 GHz........... max. 1.0 rad

Table of Contents