2.11.2 Parameter of the RS232 Interface (SYSTEM-RS232)
Submenu SYSTEM-RS232 offers access to the configuration of the RS-232 interface. The pin
assignment of the interface corresponds to the pin assignment of a PC.
Menu selection: UTILITIES - SYSTEM - RS232
FREQUENCY
LEVEL
MODULATION
DIGITAL MOD
LF OUTPUT
SWEEP
LIST
MEM SEQ
UTILITIES
HELP
SYSTEM
REF OSC
PHASE
PROTECT
CALIB
DIAG
TEST
MOD KEY
AUX I/O
BEEPER
INSTALL
DATA FORMAT
8 Bits
PARITY NO
STOP BIT 1
BAUD RATE 9600 bps
HANDSHAKE OFF RTS/CTS XON/XOFF
FREQ
100. 000 000 0
MHz
LEVEL
-30.0
dBm
GPIB
RS232
SECURITY
LANGUAGE
Fig. 2-64 Menu UTILITIES - SYSTEM - RS232
DATA FORMAT
Indication of the number of data bits. This value cannot be changed.
PARITY
Indication of parity. This value cannot be changed.
STOP BIT
Indication of the number of stop bits. This value cannot be changed.
BAUD RATE
Selection of the baud rate.
IEC-bus short command :SYST:COMM:SER:BAUD 9600
HANDSHAKE
Selection of the handshake.
OFF No handshake
IEC-bus short command :SYST:COMM:SER:PACE NONE
:SYST:COMM:SER:CONT:RTS ON
RTS/CTS Hardware handshake using the interface lines RTS and CTS. This
mode always is to be preferred to XON/XOFF mode, if permitted by
the configuration of the host computer.
IEC-bus short command :SYST:COMM:SER:CONT:RTS RFR
XON/XOFF Software handshake using the ASCII codes 11h <XON> and 13h
<XOFF>. This mode is not recommended for binary data and for
baud rates above 9600 baud.
IEC-bus short command :SYST:COMM:SER:PACE XON