Test Procedure SME
1038.6002.02 5.20 E-13
5.2.16 Modulation Depth of AM
Test equipment Test system 5.1.1.1
Test system 5.1.1.6 (SME06)
Measurement
À Settings at the SME
- Level 0 dBm
- Select LFGEN1 in the MODULATION/AM/AM SOURCE INT
menu,
- Modulation depth 0.1% to 80%
(recommended values 30% and 80%)
- Modulation frequency 1 kHz
À Vary the carrier frequency from 5 MHz to 1500 MHz (2200/
3000/6000 MHz with SME03E/03/06). For recommended setting
values, see table 5-2.
À Read off the modulation depth at the modulation analyzer.
Deviation of the modulation depth measured from the setting value
5.2.17 AM Frequency Response
Test equipment
Test system 5.1.1.2
Test system 5.1.1.6 (SME06)
Measurement
À Settings at the SME
- Level 0 dBm,
- Select EXT1 in the MODULATION/AM/AM SOURCE menu,
- Modulation depth 60%.
À Vary the carrier frequency from 5kHz to 1500MHz (2200/
3000/6000MHz with SME03E/03/06). For recommended setting
values, see table 5-2.
À Setting at the audio analyzer
- Generator level 1 V(u
peak
).
À Determine the modulation frequency response (difference between
the highest and the lowest modulation depth) by varying the
generator frequency.
Modulation frequency response from 20 Hz to 50 kHz .......max.1dB
Note: If option SM-B2 LF generator is fitted, LFGEN2 can be selected in the MODULATION/AM/AM
SOURCE menu and the test frequency be set using the internal generator for this
measurement.