SME STATus
1038.6002.02 3.115 E-13
3.6.13 STATus-System
This system contains the commands for the status reporting system (c.f. Section "Status Reporting
System"). *RST has no influence on the status registers.
Command Parameter Default
Unit
Remark
:STATus
:OPERation
[:EVENt]?
:CONDition?
:PTRansition
:NTRansition
:ENABle
:PRESet
:QUEStionable
[:EVENt]?
:CONDition?
:PTRansition
:NTRansition
:ENABle
:QUEue
[:NEXT]?
0 to 32767
0 to 32767
0 to 32767
0 to 32767
0 to 32767
0 to 32767
Query only
Query only
No query
Query only
Query only
Query only
:STATus:OPERation
The commands for the STATus:OPERation register are under this node.
:STATus:OPERation[:EVENt]?
The command queries the content of the EVENt part of the STATus:OPERation register. In
reading out, the content of the EVENt part is deleted.
Example: :STAT:OPER:EVEN? Response: 17
:STATus:OPERation:CONDition?
The command queries the content of the CONDition part of the STATus:OPERation register. In
reading out, the content of the CONDition part is not deleted. The value returned directly reflects
the current hardware state.
Example: :STAT:OPER:COND? Response: 1
:STATus:OPERation:PTRansition 0 to 32767
The command (Positive TR
ansition) sets the edge detectors of all bits of the STATus:OPERation
register from 0 to1 for the transitions of the CONDition bits.
Example: :STAT:OPER:PTR 32767
:STATus:OPERation:NTRansition 0 to 32767
The command (Negative TR
ansition) sets the edge detectors of all bits of the STATus:OPERation
register from 1 to 0 for the transitions of the CONDition bit.