SOURce:VOR SME
1038.6002.02 3.106 E-13
3.6.11.22 SOURce:VOR Subsystem
This subsystem contains the commands to control the characteristics of the test signals for VOR
(VHF O
mnidirectional Range) (option SM-B6, multifunction generator).
Command Parameter Default
Unit
Remark
[:SOURce]
:VOR
:STATe
:SOURce
:MODE
[:BANGle]
:DIRection
:VAR
[:DEPTh]
:FREQuency
:SUBCarrier
:DEPTh
[:FREQuency]
:REFerence
[:DEViation]
:PRESet
:COMid
[:STATe]
:FREQuency
:DEPTh
ON | OFF
INT2 | INT2, EXT
NORM | VAR | SUBCarrier | FMSubcarrier
0 to 360 deg
FROM | TO
0 to 100 PCT
20 to 40 Hz
0 to 100 PCT
5 to 15 kHz
0 to 960 Hz
ON | OFF
0.1 to 20 000 Hz
0 to 100 PCT
rad
PCT
Hz
PCT
Hz
Hz
Hz
PCT
Option SM-B6
No query
[:SOURce]:VOR:STATe ON | OFF
The command switches on or off the generation of VOR signals. STATe ON is only possible if no
amplitude modulation is switched on. Modulation generator 2 must not be switched simultaneously
as a source for PM or FM either. *RST value is OFF
Example: :SOUR:VOR:STAT ON
[:SOURce]:VOR:SOURce INT2 | INT2, EXT
This command determines the signal sources of the test signal.
INT2 The test signal used for VOR is generated internally by LF generator 2 (SOURce
INT2).
INT2 , EXT A signal from input EXT1 is added to the internal signal in addition. Switching off the
internal source is not possible. *RST value is INT2
Example: :SOUR:VOR:SOUR INT2