SME List of Error Messages
1038.6002.02 7B.5 E-11
SME-Specific Error Messages
Device-dependent Error - device-specific error; sets bit 3 in the ESR register.
Error code Error text in the case of queue poll
Error explanation
105
Frequency underrange
The frequency is below the limit value guaranteed.
106
Frequency overrange
The frequency is beyond the limit value guaranteed.
110
Output unleveled
The level control loop is deactivated.
115
Level overrange
The level is above the limit value guaranteed.
116
Level underrange
The level is below the limit value guaranteed.
117
Dynamic level range exceeded
The difference between the maximal and minimal value of a level list is above 20 dBm. An exact level
setting is no longer guaranteed.
120
Signal output /ERMES setting mismatch - use RECALCULATE ➤
ERMES settings have been changed during operating mode ERMES. However, the signal output is still
based on the old settings. The output signal is recalculated from the current settings by selecting
RECALCULATE ➤ in manual control or ERMES:STATe OFF; STATe ON via IEC bus.
121
ERMES: Channel/Frequency mismatch
The RF has been readjusted and does no longer match the frequency indicated in the ERMES menu
under CHANNEL.
Setting ERMES CHANNEL does not only determine the RF but also influences the structure of the
ERMES message. Thus the channel cannot simply be changed by readjusting the RF but only via
ERMES CHANNEL and RECALCULATE ➤.
125
Signal output /FLEX setting mismatch - use RECALCULATE ➤
FLEX settings have been changed during operating mode FLEX. However, the signal output is still based
on the old settings. The output signal is recalculated from the current settings by selecting
RECALCULATE ➤ in manual control or FLEX:STATe OFF; STATe ON via IEC bus.
130
FM modulator VCO unlocked
FM modulator VCO is not synchronized.
131
AM modulation frequency out of range
The AM modulation frequency is out of the permissible range.
132
PM modulation frequency out of range
The PM modulation frequency is out of the permissible range.
140
This modulation forces other modulations OFF
A modulation has been switched on which cannot be used at the same time as an already active
modulation. The previous modulation has been switched off.
152
Input voltage out of range; EXT1 too high
The input voltage at the EXT1 socket is too high.
153
Input voltage out of range; EXT1 too low
The input voltage at the EXT1 socket is too low.