List of Error Messages SME
1038.6002.02 7B.4 E-11
Device Specific Error - sets bit 3 in the ESR register.
Error code Error text in the case of queue poll
Error explanation
-310
System error
This error message suggests an error within the instrument. Please inform the R&S Service.
-311
Memory error
Error in the instrument memory.
-313
Calibration memory lost
Loss of calibration data stored. The calibration data of VCO SUM, LEVEL PRESET and PULSE GEN
can be restored by internal routines (see Chapter 2, Section Calibration).
-314
Save/recall memory lost
Loss of the non-volatile data stored using *SAV? command.
-315
Configuration memory lost
Loss of the non-volatile configuration data stored by the instrument.
-330
Self-test failed
The selftest could not be executed.
-350
Queue overflow
This error code is entered into the queue instead of the actual error code if the queue is full. It indicates
that an error has occurred but not been accepted. The queue can accept 5 entries.
-360
Communication error
An error has occurred during the transmission or reception of data on the IEC/IEEE bus or via the
RS-232 interface.
Query Error - Error in data request; sets bit 2 in the ESR register.
Error code Error text in the case of queue poll
Error explanation
-410
Query INTERRUPTED
The query has been interrupted.
Example: After a query, the instrument receives new data before the response has been sent completely.
-420
Query UNTERMINATED
The query is incomplete.
Example: The instrument is addressed as a talker and receives incomplete data.
-430
Query DEADLOCKED
The query cannot be processed.
Example: The input and output buffers are full, the instrument cannot continue operation.