Sweep SME
1038.6002.02 2.138 E-13
2.8.6 RF-Sweep
Menu SWEEP - FREQ offers access to settings for RF sweep.
Menu selection: SWEEP - FREQ
FREQUENCY
LEVEL
MODULATION
DIGITAL MOD
LF OUTPUT
SWEEP
LIST
MEM SEQ
UTILITIES
HELP
FREQ
LEVEL
LFGEN
LEVEL dBm
START FREQ
100.000 000 0 MHz
STOP FREQ 500.000 000 0 MHz
CENTER FREQ 300.000 000 0 MHz
SPAN 400.000 000 0 MHZ
CURRENT FREQ 100.000 000 0 MHz
SPACING LIN LOG
STEP LIN 1.000 000 0 MHz
DWELL 15.0 ms
MODE OFF AUTO SINGLE STEP EXT-SINGLE EXT-STEP
EXECUTE SINGLE SWEEP
RESET SWEEP
MARKER 1 FREQ 100.000 0 MHz
MARKER 1 STATE OFF ON
AMPLITUDE MARKER 1 OFF ON
MARKER 2 FREQ 200.000 0 MHz
MARKER 2 STATE OFF ON
AMPLITUDE MARKER 2 OFF ON
MARKER 3 FREQ 300.000 0 MHz
MARKER 3 STATE OFF ON
AMPLITUDE MARKER 3 OFF ON
- 30.0
START FREQ
100.000 000 0
MHz
STOP FREQ
500.000 000 0
MHz
RF-SWP
Fig. 2-55 Menu SWEEP - FREQ
2.8.6.1.1.1.1.1.1
START FREQ
Input value of the starting frequency.
IEC bus short command :SOUR:FREQ:STAR 100MHz
STOP FREQ
Input value of the stop frequency.
IEC bus short command :SOUR:FREQ:STOP 500MHz
CENTER FREQ
Input value of the center frequency.
IEC bus short command :SOUR:FREQ:CENT 300MHz
SPAN
Input value of the span.
IEC bus short command :SOUR:FREQ:SPAN 100MHz
CURRENT FREQ
Indication of the current frequency value.
Operating mode STEP: Input value of the frequency.
STEP LIN (LOG)
Input value of the step width. Depending on whether SPACING LIN or LOG
is selected, STEP LIN or STEP LOG is displayed.
IEC bus short command :SOUR:SWE:STEP:LIN 1MHz