EasyManua.ls Logo

Rohde & Schwarz SME02 - Abort System

Rohde & Schwarz SME02
462 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
SME ABORt
1038.6002.02 3.17 E-13
3.6.3 ABORt System
The ABORt system contains the commands to abort actions triggered. After an action has been
aborted, it can be triggered again at once. All commands trigger an event, thus they have no *RST
value.
Further commands for the trigger system of the SME can be found in the TRIGger system.
Command Parameter Default
Unit
Remark
:ABORt
[:SWEep]
:LIST
:MSEQuence
:DM
:XMEM
No query
No query
No query
No query /
option SME-B11
No query /
option SME-B12
:ABORt[:SWEep]
The command aborts a sweep.
Example: :ABOR:SWE
:ABORt:LIST
The command aborts a list execution.
Example: :ABOR:LIST
:ABORt:MSEQuence
The command aborts a Memory Sequence.
Example: :ABOR:MSEQ
:ABORt:DM
The command aborts the single execution of a DM list. The command always relates to the digital
type of modulation currently set using SOURce:DM:TYPE .
Example: :ABOR:DM
:ABORt:XMEM
The command aborts the recording of data into the DM memory extension, list "XMEM".
Example: :ABOR:XMEM

Table of Contents

Related product manuals