RF Level SME
1038.6002.02 2.48 E-13
STATE
Switching on/off user correction.
IEC-bus command SOUR:CORR ON
SELECT LIST...
Selection of a list or generation of a new list(cf. Section 2.2.4, List Editor)
IEC-bus command SOUR:CORR:CSET "UCOR1"
DELETE LIST...
Deletion of a list (cf. Section 2.2.4, List Editor)
IEC-bus command SOUR:CORR:CSET:DEL "UCOR2"
FUNCTION
Selection of the editing mode to process the selected list
IEC-bus commands SOUR:CORR:CSET:DATA:FREQ 100 MHz, 102
MHz,...
SOUR:CORR:CSET:DATA:POW 1dB, 0.8dB,...
FREQUENCY
LEVEL
MODULATION
DIGITAL MOD
LF OUTPUT
SWEEP
LIST
MEM SEQ
UTILITIES
HELP
LEVEL
ALC
UCOR
EMF
FREQ
100. 000 000 0
MHz
LEVEL
- 27.0
dBm
UCOR
+ 1 .9
dB
SELECT LIST... CURRENT: UCOR1
FUNCTION FILL INSERT DELETE EDIT/VIEW
INDEX - FREE 70 - LEN 10 FREQUENCY UCOR1
005 105.000 000 0 MHz + 1.9 dB
006 107.000 000 0 MHz + 1.2 dB
007 108.000 000 0 MHz + 1.3 dB
008 109.000 000 0 MHz + 1.5 dB
009 111.000 000 0 MHz + 1.6 dB
010 112.000 000 0 MHz + 1.9 dB
011 113.000 000 0 MHz + 2.0 dB
012 114.000 000 0 MHz + 2.1 dB
Fig. 2-22 Menu UCOR - LEVEL-EDIT side