DISPlay SME
1038.6002.02 3.24 E-13
3.6.6 DISPLAY-System
This system contains the commands to configure the screen. If system security is activated using
command SYSTem:SECurity ON, the display cannot be switched on and off arbitrarily (cf. below)
Command Parameter Default
Unit
Remark
:DISPlay
:ANNotation
[:ALL]
:AMPLitude
:FREQuency
ON | OFF
ON | OFF
ON | OFF
:DISPlay:ANNotation
The commands determining whether frequency and amplitude are indicated are under this node.
Caution: With SYSTem:SECurity ON, the indications cannot be switched from OFF to ON. In
this case *RST does not influence the ANNotation settings either. With
SYSTem:SECurity OFF, the *RST value is ON for all ANNotation parameters.
:DISPlay:ANNotation[:ALL] ON | OFF
The command switches the frequency and amplitude indication on or off.
Command :DISPlay:ANNotation:ALL ON can only be executed if SYSTem:SECurity is
set to OFF.
With SECurity OFF - *RST value is ON.
Example: :DISP:ANN:ALL ON
:DISPlay:ANNotation:AMPLitude
ON | OFF
The command switches on or off the amplitude indication.
Command
:DISPlay:ANNotation:AMPLitude ON can only be executed if
SYSTem:SECurity is set to OFF.
With SYSTem:SECurity OFF - *RST value is ON.
Example: :DISP:ANN:AMPL ON
:DISPlay:ANNotation:FREQuency ON | OFF
The command switches on or off the amplitude indication.
Command :DISPlay:ANNotation:AMPLitude ON can only be executed if
SYSTem:SECurity is set to OFF.
With SYSTem:SECurity OFF - *RST value is ON.
Example: :DISP:ANN:FREQ ON