SME Fast-Restore-Modus
1038.6002.02 3.139 E-13
3.8.3.5 STATus:QUEStionable Register
This register contains information on questionable instrument states. They can occur, e.g. if the
instrument is operated out of its specifications. It can be queried using commands "STATus
:QUEStionable:CONDition?" or "STATus:QUEStionable[:EVENt]?".
Table 3-7 Meaning of the bits used in the STATus:QUEStionable register
Bit-No. Meaning
0
VOLTage
This bit is set if the voltage at an output connector is not correct,
if the voltage is above or below the specified limit values,
if the level limit has responded,
if the overvoltage protection has responded.
5
FREQuency
The bit is set if a frequency at the RF output is not correct or if it is lower or higher than the specified values
7
MODulation
The bit is set if a modulation is not correct or is operated outside the specifications.
8
CALibration
The bit is set if a calibration is not performed properly.