Digital Modulation SME
1038.6002.02 2.110 E-13
NID
The following four commands define the "Network Identification" emitted
by the SME.
NETWORK
ADDRESS
Input value of the network address part of the network ID.
IEC-bus command: :SOUR:FLEX:SINF:NADD 2058240
MULTIPLIER
Input value of the multiplier part of the network ID.
IEC-bus command: :SOUR:FLEX:SINF:MULT 1
SERVICE AREA
Input value of the service area part of the network ID.
IEC-bus command: :SOUR:FLEX:SINF:SAR 0
NID TMF
Input value of the four traffic management fags for NID
IEC-bus command: :SOUR:FLEX:SINF:NTMF 15
DATE
Input value of the date.
IEC-bus command: :SOUR:FLEX:SI:DATE 1994,12,01
TIME
Input value of the current time.
IEC-bus command: :SOUR:FLEX:SI:TIME 12,00
--- MESSAGE ---
The parameters of this section are used for setting the destination
address as well as the useful data of the message.
CAPCODE
Input value of the CAPCODE of the pager to be called as printed on the
receiver. CAPCODE is defined in FLEX standard. The CAPCODE
contains the addresses of the receiver as well as frame and phase
information.
IEC-bus command: :SOUR:FLEX:MESS:CAPC "A0000001"
AUTO ADJUST
Selection of which settings are influenced by a change of CAPCODE:
NONE No influence
PHASE The phase is adjusted to the value contained in CAPCODE. This
is performed whenever the setting for CAPCODE is changed.
FRAME FRAME CONTENT is set so that in all the frames evaluated by
the receiver (and only in those) FLEX data are also transmitted.
However, FRAME CONTENTS is changed only for frames
where FRAME CONTENTS contains an ’X’, an ’x’ or a ’ ’ (blank).
For all other frames, FRAME CONTENTS is left unchanged.
BOTH PHASE as well as FRAME CONTENT are adjusted.
IEC-bus command: :SOUR:FLEX:PHAS:AUTO ON
:SOUR:FLEX:FCON:AUTO ON

Table of Contents