SME Digital Modulation
1038.6002.02 2.121 E-13
ALPHANUM MESSAGE
Opens a window to select an alphanumeric message. The
following selections can be made:
FOX „The quick brown fox jumps over the lazy dog“
ALPHA „ABCD...“ (complete ReFLEX25 character set)
USER1...4 Four messages that can be freely edited by
command EDIT MESSAGE.
There is one common set of messages for
ReFLEX25 and Flex - each modification for
ReFLEX25 also changes the message for FLEX.
IEC-bus command: SOUR:REFL25:MESS:ALPH “FOX
EDIT MESSAGE...
Opens a window to edit the message selected under ALPHANUM
MESSAGE. The maximum length of the message is 128
characters.
IEC-bus command:
SOUR :REFL25:MESS:ALPH:DATA “Hello, world!“
RESPONSE REQUIRED:
Determines whether the receiver is to confirm the message or not.
The respective bit is set in the message. In the case of OFF the
vector field is reduced by one code word if this is permitted by the
ReFLEX standard.
IEC-bus command: SOUR:REFL25:MESS:RREQ OFF
-- SYSTEM INFORMATION --
The system information parameters are used for setting the data
of the sending system (simulated by the SME). These data are
sent to the receiver in the respective block information words.
ZONE
Input value of the geographical transmit zone. The significance is
determined by the network operator. The value 0 is referred to as
“restricted“ in the ReFLEX25 documentation, but can be set for
testing. The upper limit is 4091.
IEC-bus command: SOUR:REFL25:SI:ZONE 1
SUBZONE
Input value of the subzone within a transmit zone. The
significance is determined by the network operator. The value 0 is
referred to as “restricted“ in the ReFLEX25 documentation, but
can be set for testing. The upper limit is 127.
IEC-bus command: :REFL25:SI:SZONe 1
SERVICE PROVIDER
Input value for the ID of the service provider. Valid values are 0 to
16383.
IEC-bus command: SOUR:REFL25:SI:PROV 0