SME SOURce:FREQuency
1038.6002.02 3.67 E-133
[:SOURce]:FREQuency:STOP 5 kHz to 1.5 GHz (SME03E/03/06: 5 kHz to 2.2/3/6 GHz)
This command indicates the final value of the frequency for the sweep operation (see STARt as
well). (As to specified range, cf. FREQuency:CENTer). *RST value is 500MHz
Example: :SOUR:FREQ:STOP 100MHz
[:SOURce]:FREQuency:STEP
The command to enter the step width for the frequency setting if frequency values UP or DOWN
are used is under this node. This command is coupled to the KNOB STEP command in manual
control. Only linear step widths can be set.
[:SOURce]:FREQuency:STEP[:INCRement] 0 to 1 GHz
The command sets the step width for the frequency setting.
Example: :SOUR:FREQ:STEP:INCR 1MHz *RST value is 1MHz